89
Reference Manual
00809-0200-4840, Rev BA
Section 5: Service and Troubleshooting
December 2014
Section 5: Service and Troubleshooting
5.5.6
Current output (D906)
This indicates the actual output current in mA.
Note
When the Poll Address (
) is a non-zero number, the transmitter is in
multi-drop
mode
and the current output is fixed at 4 mA.
Field Communicator or AMS
To view the actual output current:
1.
From the
Home
screen, select 3: Service Tools.
2.
Select 2: Variables.
3.
Select 2: Analog Output.
4.
Select 1: Current Output.
5.
When finished, press the Bksp key to exit to the previous menu.
Rosemount 3490 Series Control Unit
To view the actual output current:
1.
From the
Main Menu
screen, select MONITOR.
2.
Select the transmitter (e.g. “
Tx1: 3107
”).
3.
Select READINGS, and then select CURRENT.
4.
Select Current Output.
5.
When finished, press the ESC key to exit to the previous menu.
Fast Keys
3, 2, 2, 1
Current Output - (mA)
Value: 19.047
HELP
0
20
15
10
5
(Field Communicator Screen)
Current Output D906
19.05mA
(Rosemount 3491 Screen)
Содержание 3107
Страница 2: ......
Страница 4: ...Reference Manual 00809 0200 4840 Rev BA Title Page December 2014 ...
Страница 10: ...TOC vi Reference Manual 00809 0200 4840 Rev BA Table of Contents December 2014 Table of Contents ...
Страница 16: ...6 Reference Manual 00809 0200 4840 Rev BA Section 2 Transmitter Overview December 2014 ...
Страница 90: ...80 Reference Manual 00809 0200 4840 Rev BA Section 4 Configuration December 2014 Section 4 Configuration ...
Страница 144: ...134 Reference Manual 00809 0200 4840 Rev BA Appendix D Field Communicator December 2014 Appendix D Field Communicator ...
Страница 145: ......
Страница 146: ......
Страница 147: ......