97
Reference Manual
00809-0200-4840, Rev BA
Section 5: Service and Troubleshooting
December 2014
Section 5: Service and Troubleshooting
5.5.14
Threshold in use (D917)
This indicates the echo strength threshold limit, below which echoes are rejected.
Note
The factory default setting is to automatically set the threshold level for optimum
performance (see Set Threshold on
Field Communicator or AMS
To view the threshold limit:
1.
From the
Home
screen, select 3: Service Tools.
2.
Select 1: Alerts.
3.
Select 2: Diagnostic Data.
4.
Select 3: Operation.
5.
Select 2: Threshold In Use D917.
6.
When finished, press EXIT to exit to the previous menu.
Note
This is also available at the Fast Key sequence 2, 2, 7, 2, 4
(see
Appendix D: Field Communicator
for a full menu map).
Rosemount 3490 Series Control Unit
To view the threshold limit:
1.
From the
Main Menu
screen, select MONITOR.
2.
Select the transmitter (e.g. “
Tx1: 3107
”).
3.
Select DIAGNOSTICS.
4.
Select OPERATION.
5.
Select Thresh In Use.
6.
When finished, press the ESC key to exit to the previous menu.
Fast Keys
3, 1, 2, 3, 2
Threshold in Use D917
12 %
EXIT
HELP
(Field Communicator Screen)
Thresh In Use D917
12%
(Rosemount 3491 Screen)
Содержание 3107
Страница 2: ......
Страница 4: ...Reference Manual 00809 0200 4840 Rev BA Title Page December 2014 ...
Страница 10: ...TOC vi Reference Manual 00809 0200 4840 Rev BA Table of Contents December 2014 Table of Contents ...
Страница 16: ...6 Reference Manual 00809 0200 4840 Rev BA Section 2 Transmitter Overview December 2014 ...
Страница 90: ...80 Reference Manual 00809 0200 4840 Rev BA Section 4 Configuration December 2014 Section 4 Configuration ...
Страница 144: ...134 Reference Manual 00809 0200 4840 Rev BA Appendix D Field Communicator December 2014 Appendix D Field Communicator ...
Страница 145: ......
Страница 146: ......
Страница 147: ......