27
Reference Manual
00809-0200-4840, Rev BA
December 2014
Section 4: Configuration
4.2.7
Restart device
This re-starts the transmitter as if the power has been interrupted.
Field Communicator or AMS
To restore the original factory configuration:
1.
From the
Home
screen, select
3: Service Tools
.
2.
Select
4: Maintenance
.
3.
Select
3: Utilities
.
4.
Select
1: Restart Device
.
5.
When the message “About to restart the transmitter” appears, press “
OK
” to continue
(or “
ABORT
” to not continue).
6.
Select
1: Yes
to restart the transmitter.
Note
When messages appear, take appropriate action if needed and press “
OK
”.
Rosemount 3490 Series Control Unit
To restore the original factory configuration:
1.
From the
Main Menu
screen, select
SETUP
.
2.
Select the transmitter (e.g. “
Tx1: 3107
”).
3.
Select
SYSTEM
, and then select
RESTART DEVICE
.
4.
Follow the on-screen instruction (“
Start
”) to restart the transmitter.
(Press the
Enter
(
) key if prompted to change the mode to “off-line”).
5.
Select “
Quit
” to exit to the previous menu.
Fast Keys
3, 4, 3, 1
Utilities
1. Restart Device
2. Load Defaults
3. Set Base Units
1. Restart Device
SAVE
HOME
HELP
(Field Communicator Screen)
Esc=Quit =Start
RESTART DEVICE
(Rosemount 3491 Screen)
Содержание 3107
Страница 2: ......
Страница 4: ...Reference Manual 00809 0200 4840 Rev BA Title Page December 2014 ...
Страница 10: ...TOC vi Reference Manual 00809 0200 4840 Rev BA Table of Contents December 2014 Table of Contents ...
Страница 16: ...6 Reference Manual 00809 0200 4840 Rev BA Section 2 Transmitter Overview December 2014 ...
Страница 90: ...80 Reference Manual 00809 0200 4840 Rev BA Section 4 Configuration December 2014 Section 4 Configuration ...
Страница 144: ...134 Reference Manual 00809 0200 4840 Rev BA Appendix D Field Communicator December 2014 Appendix D Field Communicator ...
Страница 145: ......
Страница 146: ......