70
Guidance and manufacturers declaration - electromagnetic immunity
The Ultimate Comfort is intended for use in the electromagnetic environment specified below. The customer or the user of the Ultimate Comfort should assure that
it is used in such an environment.
Immunity tests
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
according IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
according IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a typical commer-
cial or hospital environment.
Surge
according IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical commer-
cial or hospital environment.
Interruptions and voltage varia-
tions on power supply input
lines
according to IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
for 0.5 cycle)
40 % U
T
(60 % dip in U
T
for 5 cycles)
70 % U
T
(30 % dip in U
T
for 25 cycles)
<5 % U
T
(>95 % dip in U
T
for 5 s)
<5 % U
T
(>95 % dip in U
T
for 0.5 cycle)
40 % U
T
(60 % dip in U
T
for 5 cycles)
70 % U
T
(30 % dip in U
T
for 25 cycles)
<5 % U
T
(>95 % dip in U
T
for 5 s)
Mains power quality should be that of a typical commer-
cial or hospital environment. If the user of the Ultimate
Comfort requires continued operation during power
mains interruptions, it is recommended that the
WED-3000 B Ultrasound Scanner be powered from an
uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic filed
according IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
REMARK
U
T
is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturers declaration - electromagnetic emission
The Ultimate Comfort is intended for use in the electromagnetic environment specified below. The customer or the user of the model Ultimate Comfort should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
according CISPR 11
Group 1
The Ultimate Comfort uses RF energy only for its internal function. Therefor, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
according CISPR 11
Class B
The Ultimate Comfort system is suitable for use in all establishments, inclu-
ding domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
according IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
according IEC 61000-3-3
Compliant