9 Troubleshooting
9-1. <FAILURE>
70 / 87
Display
State explanation
Main cause and required action
Main controller:
Abnormal temperature
detected
Unless an abnormality is found in the installation
environment, a malfunction of the main controller
is suspected.
Main controller:
Supply power abnormality
detected
Unless an abnormality is found in the OHC-800
supply power, a malfunction of the power
terminal unit or main controller is suspected.
Main controller:
Optical sensor unit
abnormality detected
A malfunction of the connecting cable between
the main controller and the optical sensor unit or
the optical sensor unit is suspected.
Main controller:
Sonic sensor unit
abnormality detected
A malfunction of the connecting cable between
the main controller and the sonic sensor unit or
the sonic sensor unit is suspected.
Main controller:
4 - 20 mA output signal
abnormality detected
Unless an abnormality such as a break in 4 - 20
mA cable is found, a malfunction of the power
terminal unit or main controller is suspected.
Optical sensor unit:
ROM abnormality detected
An abnormality in the optical sensor unit ROM.
The optical sensor unit must be replaced.
Optical sensor unit:
RAM abnormality detected
An abnormality in the optical sensor unit RAM.
The optical sensor unit must be replaced.
Optical sensor unit:
FRAM abnormality detected
An abnormality in the optical sensor unit FRAM.
The optical sensor unit must be replaced.
Optical sensor unit:
Temperature abnormality
detected
Unless an abnormality is found in the installation
environment, a malfunction of the optical sensor
unit is suspected.
Optical sensor unit:
GAS OUT abnormal
absolute pressure detected
Unless an abnormality in the GAS OUT exhaust
environment or tubing such as clogging is found,
a malfunction of the optical sensor unit is
suspected.
Optical sensor unit:
Supply power abnormality
detected
Unless an abnormality is found in the OHC-800
supply power, a malfunction of the power
terminal unit or optical sensor unit is suspected.
Optical sensor unit:
Abnormally low brightness
of interference fringe image
data detected
A damage of the optical sensor unit due to
inhalation of foreign substances or waste
materials is suspected.
Содержание OHC-800
Страница 1: ...PT3E 04914 Explosion Proof Calorimeter OHC 800 Operating Manual...
Страница 88: ......