5 How to Operate in Check Mode
5-2. Items and details
47 / 87
5-2-13. Latching reset for display and contact (LATCHING
RESET (DISP. & CONTACT))
FAILURE
OUT OF SPECIFICATION
MAINTENANCE REQUIRED
If the OHC-800 detects one of the state categories shown above using the self-diagnostic/monitoring
function and returns to a normal state by itself, the contact is self-latched and the LCD display screen enters
the "Trace display" state in its standard settings. (See Section 4.5, "Operations of contacts, display, and
signal output after recovery to normal state.")
Every 1 second
Operation of "Trace display"
This menu item is used to reset the self-latch state of the contacts and the "Trace display" state of the LCD
display.
In the menu screen, press the SET key to display the CAUTION screen informing that the latching states of
the display screen and contacts will be reset.
When "OK" is selected using the
▲▼
+ SET keys, the states of the contacts and LCD are reset, and the
ordinary measurement screen and the normal contact operation are resumed.
Menu screen
CAUTION screen
(English)
CAUTION screen
(Chinese)
CAUTION screen
(Japanese)