5 How to Operate in Check Mode
5-2. Items and details
44 / 87
5-2-7. Check of pressure sensor output (PRESSURE SENSOR
READINGS)
This screen displays the output of three pressure sensors installed in the optical sensor unit.
P (GAS): Output of the micro-differential pressure sensor that detects the flow rate of measuring
gas supplied to the OHC-800.
P (REF): Output of the micro-differential pressure sensor that detects the flow rate of REF gas
supplied to the OHC-800.
P (OUT): Output of the absolute pressure sensor at GAS OUT of the OHC-800 to be used for
pressure correction.
Output value of GAS differential pressure
sensor
Output value of REF differential pressure
sensor
Output value of OUT absolute pressure
sensor
Menu screen
5-2-8. Check of temperature sensor output (TEMPERATURE
SENSOR READINGS)
This screen displays the output of temperature sensors installed in the main controller, optical sensor unit,
and sonic sensor unit.
Output value of temperature sensor on
main PCB
Output value of temperature sensor on
optical sensor unit
Output value of temperature sensor on
sonic sensor unit
Menu screen