6 How to Operate in SETUP MODE
6-2. Items and details
53 / 87
6-2-3. 4 – 20 mA condition settings (4 – 20 mA SETTINGS)
Display the output conditions of 4 - 20 mA signal. In the menu screen, press the SET key to display the
setting value check screen and display the current settings of 4 - 20 mA conditions.
Menu screen
Setting value check
screen
The example above shows the screens displayed when "CALORIFIC VALUE" is the selected output item,
and the 36.00 to 46.00 MJ/m
3
range is the output range setting of 4 - 20 mA signals.
When the SET key is pressed in the setting value check screen, the "CAUTION screen" is displayed to
inform that 4 - 20 mA output signal will be changed.
CAUTION screen
(English)
CAUTION screen
(Chinese)
CAUTION screen
(Japanese)
When "OK" is selected using the
▲▼
+SET keys, and the setting screen appears. As shown in the figure
below, use the
▲▼
and SET keys to specify the output item, upper and lower limits of the output range, and
output values for the FAILURE and FUNCTION CHECK states.
Setting of output item
Setting of lower limit of
output range
Setting of upper limit of
output range
Setting of output value
for FAILURE
Setting of output value for
FUNCTION CHECK