3 How to Install Calorimeter
3-3. How to connect wire
23 / 87
3-3-2. Recommended cables
Connected to
Recommended cable
Cable overall outer diameter
Power (AC) line
CVV 1.25 mm
2
/3-core
Φ
10.0
CVV 2 mm
2
/3-core
Φ
11.0
Power (DC) line
CVVS 1.25 mm
2
/2-core
Φ
10.0
CVVS 2 mm
2
/2-core
Φ
11.0
4 - 20 mA line
CVVS 1.25 mm
2
/2-core
Φ
10.0
CVVS 2 mm
2
/2-core
Φ
11.0
Contact x 1 line
CVVS 1.25 mm
2
/2-core
Φ
10.0
CVVS 2 mm
2
/2-core
Φ
11.0
Contact x 2 lines
CVVS 1.25 mm
2
/4-core
Φ
11.0
CVVS 2 mm
2
/4-core
Φ
12.0
Contact x 3 lines
CVVS 1.25 mm
2
/6-core
Φ
13.0
CVVS 2 mm
2
/6-core
Φ
14.0
RS-485 line
KPEVS 0.75 mm
2
/2P
Φ
11.0
CVV = Control -use Vinyl insulated Vinyl sheathed cable
CVVS = Control -use Vinyl insulated Vinyl sheathed cable with shield
NOTE
The overall outer diameters must be checked because they may somewhat vary between manufacturers.