6 How to Operate in SETUP MODE
6-1. SETUP MODE items
50 / 87
6-1. SETUP MODE items
In the SETUP MODE, the menu items shown in the table below can be used.
Setting of calorific value
calculation conditions
Setting of density calculation
conditions
Setting of 4 - 20 mA conditions
Set the detailed conditions for
calorific value calculation such as
sonic sensor measurement,
optical sensor measurement, and
Opt-Sonic calculation.
Set the detailed conditions for
density calculation such as sonic
sensor measurement, optical
sensor measurement, and
Opt-Sonic calculation.
Set the details of information to
be output in 4 - 20 mA signals.
4 - 20 mA output adjustment
4 - 20 mA output test
Reference calibration
Adjust 4 - 20 mA signal output
level.
Output an arbitrary test signal for
4 - 20 mA signals.
Perform reference calibration.
Reference calibration is
operation for adjusting the
reference point of the optical
sensor unit.
Offset
Display/contact operation
settings
Change of LCD display settings
Adjust the offset value to be
added to or subtracted from the
measuring results of calorific
value or density.
Set the details of display and
contact operations when
detecting a certain condition
using the self-diagnostic function.
Set the contrast of LCD display
and brightness of backlight.
Change of RS-485 (MODBUS)
communication settings
Change of energized contact
settings
Check the contact operation
Set the RS-485 (MODBUS)
communication conditions.
Set the energized contact
conditions for Contact Outputs 1
through 3.
Output an arbitrary test signal for
contact signal.
Password change
Log data download
Change the password required to
enter SETUP MODE.
Download the log data using
IrDA.