5 How to Operate in Check Mode
5-2. Items and details
46 / 87
5-2-11. Check of Opt-Sonic calculation process (OPT-SONIC
READINGS)
This screen displays the interim result of the Opt-Sonic calculation.
Menu screen
The items to be displayed are as shown below.
Symbol
Meaning of symbol
α
Factor used in the calorific value calculation of Opt-Sonic calculation
H(OPT)
Calorific value measured using only the optical sensor
H(SONIC)
Calorific value measured using only the sonic sensor
H(OUT)
Result obtained using the method selected in the "calorific value calculation setting"
β
Factor used in the density calculation of Opt-Sonic calculation
S(OPT)
Density measured using only the optical sensor (converted to specific gravity)
S(SONIC)
Density measured using only the sonic sensor (converted to specific gravity)
S(OUT)
Result obtained using the method selected in the "density calculation setting"
Regardless of the settings of the product, this screen displays a calorific value converted to MJ/m
3
, Gross,
0°C, and 101.325 kPa and a density converted to a specific gravity (AIR=1), respectively.
5-2-12. Check of display and contact settings (DISP. &
CONTACT PARAMETER)
This screen displays how the LCD display and contacts operate according to the conditions applied when
the OHC-800 detects either of the FAILURE, FUNCTION CHECK, OUT OF SPECIFICATION, and
MAINTENANCE REQUIRED states.
In the menu screen, press the SET key to display the list of conditions of the self-diagnostic/monitoring
function. Using the
▲▼
keys, select a condition and press the SET key to display the LCD display and
contact operations when the selected condition is detected.
For information on how to change the settings and available operations, see the SETUP MODE described
in Section 6.2.8, "Setting of display and contact operations (DISP. & CONTACT SETTINGS)."
The self-diagnostic conditions are displayed.
Select an item using
▲▼
keys and press the
SET key to check the details.
Menu screen
The details of LCD display and contact
operation under the self-diagnostic
conditions are displayed.