1 Important Notices on Safety
1-1. Danger cases
5 / 87
1
Important Notices on
Safety
1-1. Danger cases
DANGER
<About explosion-proof>
Satisfy the following installtaion requirements of the calorimeter.
Do not open the lid during operation.
Turn off the power of the calorimeter before replacing the fuse.
(Fuse specifications: 250 V, 1 A,
Φ
5 x 20 mm, time lag type)
Do not disassemble/modify the calorimeter, or change the settings if not necessary.
Contact RIKEN KEIKI for information on readjustment including gas calibration and parts
replacement.
Do not replace parts at your sole discretion but contact RIKEN KEIKI if the transparent window has a
crack or the explosion-proof joint surface is abnormal, or the clamping screw or bolt is changed, lost
etc.
Never fail to perform a regular maintenance.
Use appropriate cables to wire.
Use ring terminals etc. for electrical connections and ensure safe connection without any slack or
tangle.
Use ring terminals to connect the grounding terminal with the ground and use the grounding wire
with cross-sectional area of 4 mm
2
or more for the external grounding terminal.
The transparent window is made of polycarbonate resin. Be careful that it may be damaged by an
atmosphere of toluene, benzene, ammonia, aromatic hydrocarbon etc.
Do not repair the explosion-proof joint surface.
Use a control key (magnet) specified by RIKEN KEIKI.
Gas conditions (Explosion protection conditions)
Maximum flow rate
Measuring gas (GAS IN) 1 L/min
Reference gas (REF IN) 0.5 L/min
Pressure
80 to 110 kPa for all of GAS IN, REF IN, and OUT
Temperature
-20 to 57°C for all of GAS IN, REF IN, and OUT
(Japanese explosion-proof spec.)
-20 to 60°C for all of GAS IN, REF IN, and OUT
(Overseas explosion-proof spec.)