6 How to Operate in SETUP MODE
6-2. Items and details
57 / 87
6-2-7. Offset (OFFSET ADJUSTMENT)
Adjust the offset value to be added to or subtracted from the measuring results of calorific value or density.
Use this function if there is a difference between the measuring results of the OHC-800 and those of the
calorimeter/densiometer that the user uses as the criteria or those of the standard gas.
In the menu screen, press the SET key to display the setting value check screen and display the current
settings of offset values. Next, press the SET key to display the input screen and enter and confirm the
offset values for the calorific value and density in turn using the
▲▼
and SET keys.
Menu screen
Setting value check
screen
Input screen
Input screen