5 How to Operate in Check Mode
5-2. Items and details
42 / 87
5-2-2. Check of sonic sensor unit state (SONIC SENSOR UNIT
CONDITION)
This screen displays the program information for the sonic sensor unit, the self-diagnostic results measured
inside the unit etc.
Displays the program number,
SUM number, and
serial number.
Menu screen
Displays the time in which sound arrives,
standardized signal, and
self-diagnostic flags
(failure and warning flags).
5-2-3. Check of main controller state (MAIN CONTROLLER
CONDITION)
This screen displays the program information for the main controller, the self-diagnostic results measured
inside the controller etc.
Displays the program number, SUM number,
Rev number, SPE number, serial number
etc.
Menu screen
Displays the voltages of the power supply
systems (3.3 V, 5 V, and 24 V systems).
Display the output state of 4 - 20 mA signal.
Displays the various self-diagnostic flags
monitored inside the main controller.
Displays the language used for "CAUTION
screen."
Содержание OHC-800
Страница 1: ...PT3E 04914 Explosion Proof Calorimeter OHC 800 Operating Manual...
Страница 88: ......