Operative Instructions
32
10 Annex
10.1 Menu parameter table
All parameters that can be changed with the technician
are listed below.
a
WARNING! Do not change any of the
parameters with the "Only with technician"
checkbox without the support of a
service personnel or the manufacturer.
Unauthorized modification may cause
damage to the machine. The manufacturer
is not responsible for failure to follow these
instructions. The execution of unauthorised
changes, both on the machine and on the
programs, are the cause of cancellation by
the manufacturer of the warranty granted.
ID
Parameter Name
Min
Max
Description
Factory
Default
Only with
techni-
cian*
PROGRAM
1
Time phase*
0
99
Set the duration time for each
phase of the selected program
and the duration time of the
disinfection once the set
temperature is reached
NA
YES
2
Softener erogation*
0.4
1
18
45
Set the dosage of antiscale
dispensed
4.8 sec
12 ml
YES
3
Disinfection
temperature*
80
90
Define the temperature
required for items to be
considered disinfected
85 °C
YES
SET CLOCK
1
Date and time
DD/MM/YY
HH:MM:SS
Set the current date and time
NO
VARIOUS PARAMETERS
1
Delta temp. alarm*
0
20
Maximum phase temperature
deviation
15°C
YES
2
N° cycles for service*
0
9900
Set number of cycles for
maintenance request
0
YES
3
Heating timeout*
1
35
Maximum time for set
temperature reached
10 min
YES
4
Disinfection timeout*
1
30
Maximum time for set
disinfection temperature
reached
15 min
YES
Содержание DVS