To install the Cooling Fan
1. Align the cooling fan with the holes in the Compressor Cover.
2. Insert then tighten the four screws to secure the fan to the Compressor Cover.
3. Connect the fan wires to the terminals on the side of the cooling fan.
NOTE
The fan must be installed so the terminals are
oriented towards the front and center of the unit.
NOTE
Ensure that the convoluted tubing covering the fan wires is still properly seated in the
wire holder built into the front of the Compressor.