Test the condition of the 9-volt battery as follows:
1. With the unit disconnected from the power source, move the Power Switch to the ON (I) position
and verify that the red LED alarm illuminates and the audible alarm sounds.
2. The battery should provide sufficient power to alarm for 10 minutes.
3. If the red LED does not illuminate or the audible alarm does not sound, install a new 9-volt battery.
Be sure to move the Power Switch to the OFF (0) position, then proceed as follows.
a. Remove the rear access door.
b. Remove the Pre-inlet Filter, and the Inlet Filter.
c. Locate the 9-volt battery, and the battery holder clips. Remove the battery from its holder.
d. Remove the battery connector from the battery and check the no-load voltage with a DMM. If
the voltage is less than 5 volts, install a new battery.
S
YSTEM
S
ELF
T
EST
AND
S
TART
U
P
T
EST
1. Connect the Power Cord to a power source.
2. Turn on the unit by moving the Power Switch to the ON (I) position and verify the following:
•
All LEDs light up and the audible alarm sounds for two seconds.
•
The unit starts running.
•
The red LED light goes off and the audible alarm stops.
•
The yellow LED blinks until the specified oxygen level is reached (OPI version Only).
S
YSTEM
P
RESSURE
T
EST
The system pressure test is used to verify the internal operating pressures of the Millennium Oxygen
Concentrator.
1. Turn on the unit by moving the Power Switch to the ON (I) position. Let the unit run for a minimum
of one hour.
2. Turn off the Millennium Oxygen Concentrator.
3. Remove the Front and Rear Cabinets from the unit.
4. Disconnect the yellow pressure tubing from the fitting on the top of the Sieve Canister Assembly.
Using the pressure gauge, “T” fitting, and pressure tubing supplied with the Millennium tool kit
(Respironics p/n H647), connect a pressure gauge to the long yellow pressure tubing, and the
Sieve Canister Assembly.
5. Turn the unit on, and set the flow to 10 lpm. Allow the unit to cycle for at least two (2) minutes to
stabilize.
6. Hold the pressure gauge in a vertical position.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...