M
AINTENANCE
- P
AGE
21
1041813, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
21. To remove the connecting rod assembly from the shaft, rotate the eccentric until the set screw
lines up with the Compressor housing access hole. Insert the 5/32” Allen wrench or hex bit
through the access hole and loosen the set screw 1/4 turn. Rotate the eccentric to top dead center
(180
°),
and slide the connecting rod/eccentric assembly off of the shaft.
F
IGURE
Z: R
EMOVING
THE
E
CCENTRIC
22. Wrap the neck of the connecting rod assembly with a cloth, and secure the assembly in a vice
being cautious not to clamp down on the eccentric and or bearings.
F
IGURE
AA: C
ONNECTING
R
OD
A
SSEMBLY
IN
V
ICE
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...