5. Remove the two small screws which secure the Hour Meter.
F
IGURE
AF: S
CREW
L
OCATIONS
6. Remove the Hour Meter from the Front Cabinet.
To install the Hour Meter
1. Install the Hour Meter into the Front Cabinet. Hand tighten the two small screws which secure the
Hour Meter.
2. Connect the wires from the Hour Meter to terminals J16 and J17 on the Main PCB assembly.
Ensure that the wires are fully seated on the Main PCB assembly. Refer to the block diagram in
Figure AE.
3. If Overlay was not damaged when removed, install Overlay onto cabinet. If Overlay was damaged
during removal then install new Overlay onto Front Cabinet.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...