To remove the Hour Meter
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the Compressor Cover. Refer to the Compressor Cover Replacement Section.
4. Using your fingers or needle-nose pliers, remove the wires from the terminals on the back of the
Hour Meter.
5. Spread the tabs on the locking frame out from the Hour Meter.
6. While holding the tabs spread away from the Hour Meter, push it through the openings in the Base
Cabinet.
F
IGURE
AJ: R
EMOVAL
OF
THE
H
OUR
M
ETER
To install the Hour Meter
1. Align then insert the Hour Meter into its receptacle in the Base Cabinet.
2. While holding the Hour Meter in place, install the locking frame over the Hour Meter from inside of
the Base Cabinet. Press it onto the Hour Meter until it “snaps” in place.
3. Install the Hour Meter wires on the terminals on the back of the Hour Meter.
NOTE
The Hour Meter wires are interchangeable.
Remove Wires
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...