T
ROUBLESHOOTING
- P
AGE
5
1041814, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
SYMPTOM
CAUSE
VERIFICATION
CORRECTIVE ACTION
•
High pressure
alarm
•
Low pressure alarm
•
Oxygen percentage
indicator
•
Pressure relief
valve activated
Valve/Solenoid failure
•
Low oxygen
production
•
No oxygen
production
•
Supply pressure too
high
•
verify the solenoid
is working
•
Verify the pilot
valve is working
•
Verify the main
valve is working
•
Verify all wiring
harnesses are
connected
•
Verify the
Compressor is
working correctly
•
Replace the valve/
solenoid
•
Replace the valve/
solenoid
•
Replace the valve/
solenoid
•
Reconnect the wiring
harnesses
•
Rebuild the Compressor
•
Replace the Compressor
•
Low pressure alarm
Sieve Canister leaking
•
Reduced or no
oxygen production
•
Verify there are no
loose screws on
the Sieve Canister
Assembly
•
Verify there are no
blown gaskets on
the Sieve Canister
Assembly
•
Tighten the hardware
•
Replace the Sieve
Canister
Cooling fan failure
•
Warm cabinet
temperature
•
Thermal switch open
•
Verify the cooling
fan electrical
connectors are
seated
•
Verify there is
continuity on
electrical fan
wiring harness
•
Verify there is
power to the
cooling fan
•
Verify the cooling
fan bearings are
not worn
•
Verify the cooling
fan is not
obstructed
•
Reseat the connectors
•
Replace the wiring
harness Main Power
•
Reseat the connectors
•
Replace the fan assembly
•
Remove the obstructions
from the cooling fan
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...