This section describes the procedures that should be performed prior to patient use and periodically to ensure
proper setup and operation.
S
YSTEM
S
ETUP
1. Remove the rear access door and verify that the proper filters are in place.
2. Ensure that filters are secure. Twist the filter while firmly pressing downwards.
3. Before connecting the unit to a power source, perform the following procedure to check battery
condition and alarm function.
a. Move the Power Switch to the ON (I) position. The alarm should sound and the red LED
should illuminate.
b. Move the Power Switch to the OFF (0) position. The alarm and the red LED light should go
off.
4. Connect the Power Cord to a power source.
5. Turn on the unit by moving the Power Switch to the ON (I) position and verify the following:
•
all LED’s illuminate and audible alarm sounds for two seconds;
•
the unit starts running; and
•
the green power LED remains illuminated.
6. Adjust the Flow Meter to max. 5 or 10 lpm depending on the concentrator. Turning the Flow Meter
clockwise decreases the flow, and turning it counter-clockwise increases the flow.
7. If the items above are working properly, proceed to step 8. If there is a problem refer to the Trou-
bleshooting section.
8. Attach a calibrated Oxygen Analyzer to the DISS outlet fitting. The oxygen concentration should
be as specified in the
Specifications
section.
NOTE
If you have a Model 605, which is equipped with the oxygen percentage indicator (OPI), the yellow
LED light will blink until the oxygen purity is above 85% for H600/605 Concentrators and above
82% for M600/605; Enhanced M600/605 Concentrators. The green normal oxygen LED
illuminates, once appropriate O
2
purity is achieved.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...