T
ESTING
- P
AGE
3
1041903, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
10. If the regulated pressure needs adjusted, place a 5/32” Allen wrench in the adjusting screw on the
end of the regulator.
11. Disconnect the pressure gauge and “T” fitting.
12. Reconnect the clear oxygen tubing to the pressure regulator.
13. Install the Front and Rear Cabinets.
NOTE
If the regulated pressure is low, turn the wrench clockwise to increase the regulated pressure. If the
regulated pressure is high, turn the wrench counter clockwise to decrease the pressure.