E
NHANCED
M600/605; E
NHANCED
M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
17
1041897, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
OPI PCB R
EPLACEMENT
(M
ODEL
605 O
NLY
)
To remove the OPI Board
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the OPI wiring harness connector from the J7 location on the OPI board by pulling the
connector directly up from the J7 receptacle.
4. While holding the oxygen sensor, remove the pressure tubing from the sensor by carefully prying
the tubing off both oxygen sensor fittings.
5. Using a Phillips screwdriver, remove the four screws securing the OPI board to the Front Cabinet.
F
IGURE
P: C
ONNECTION
AND
M
OUNTING
S
CREW
L
OCATIONS
6. While holding the wiring harness out of the way, remove the OPI board from the Front Cabinet.
NOTE
The pressure tubing from the Flow Meter is connected to the top fitting on the oxygen sensor.
The pressure tubing connected to the pressure regulator on the Sieve Canister Assembly is
connected to the bottom fitting on the oxygen sensor.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...