OPI VOLTAGE VERIFICATION CHART
MODEL 605 ONLY
Oxygen Concentration
Percentage Average
Average Acceptable Voltage
Output (VDC)
96%
3.96 to 3.72 VDC
95%
3.92 to 3.68 VDC
94%
3.88 to 3.64 VDC
93%
3.84 to 3.60 VDC
92%
3.80 to 3.56 VDC
91%
3.76 to 3.52 VDC
90%
3.72 to 3.48 VDC
89%
3.68 to 3.44 VDC
88%
3.64 to 3.40 VDC
87%
3.60 to 3.36 VDC
86%
3.56 to 3.32 VDC
85%
3.52 to 3.28 VDC
82%
3.40 to 3.16 VDC