M10600/M10605 R
EPAIR
& R
EPLACEMENT
- P
AGE
19
1041898, R
EV
. 01
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
F
IGURE
T: R
EMOVING
THE
OPI B
OARD
To install the OPI Board
1. Align the holes in the OPI board with the standoffs on the Front Cabinet.
2. Insert then tighten the four screws provided to secure the OPI board to the Front Cabinet.
3. While holding the oxygen sensor, install the pressure tubing on the oxygen sensor by carefully
pushing the tubing onto the oxygen sensor fittings.
4. Align the connector on the OPI wiring harness with the J7 receptacle on the OPI board. carefully
press the connector onto the receptacle until completely seated.
NOTE
Once the unit is reassembled, verify that there are no O
2
leaks. Set the Flow Meter to 0 lpm. Ensure
the “no flow” alert activates. The alert should activate. If not, check for leaks at newly installed OPI
board.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...