1041813, R
EV
. 00
P
AGE
10 - M
AINTENANCE
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
10. Remove the screws, and replace the valve flappers on the valve plate using a 1/4 in. nut driver.
F
IGURE
K: F
LAPPER
V
ALVE
R
EMOVAL
11. Remove and replace the gasket on the valve plate.
12. Using a T-20 torx bit, remove and discard the four screws, the piston cup, the retainer plate and
the cork gasket from the connecting rod assembly.
NOTE
The Valve flappers are located under the keeper strips.
NOTE
A heat gun may be used to loosen the loctite on the screws.
Valve Keeper
Strips
Valve Gasket