1041895, R
EV
. 00
P
AGE
46 - H600/H605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
5. Lift the Valve/Solenoid Assembly up and away from the Sieve Canister Assembly.
F
IGURE
AT: R
EMOVING
THE
V
ALVE
/S
OLENOID
A
SSEMBLY
To install the Valve/Solenoid Assembly
1. Align the Valve/Solenoid Assembly with the holes in the top of the Sieve Canister Assembly.
2. Using the five screws and flat washers provided, secure the Valve/Solenoid Assembly to the Sieve
Canister Assembly. The screws should be torqued sufficiently in order to prevent leaks, but care
should be taken not to crush the shoulder of the valve.
3. Reconnect the two wire connectors to the Valve/Solenoid Assembly.
4. Connect the unit to a power source and turn on the device. Apply snoop leak detector to the area
where the Valve/Solenoid Assembly mounts to the Sieve Canister Assembly and check for leaks.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...