1041898, R
EV
. 01
P
AGE
40 - M10600/M10605 R
EPAIR
& R
EPLACEMENT
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
S
IEVE
C
ANISTER
A
SSEMBLY
R
EPLACEMENT
To remove the Sieve Canister Assembly
1. Remove the Rear Cabinet. Refer to the Rear Cabinet Replacement Section.
2. Slide the Front Cabinet forward until it is clear from the Base Cabinet.
3. Remove the Compressor Cover. Refer to the Compressor Cover Replacement Section.
4. Remove the Valve/Solenoid Assembly. Refer to the Valve/Solenoid Replacement Section.
5. Remove the pressure regulator. Refer to the Pressure Regulator Replacement Section.
6. Using a phillips screwdriver, remove the four screws securing the Sieve Canister Assembly to the
Base Cabinet.
7. Using a crimping clamp tool or diagonals, cut the one-eared clamp securing the 12” pressure tub-
ing to the inlet port of the Sieve Canister.
8. Remove the 12” pressure tubing from the inlet port of the Sieve Canister.
9. Lift the Sieve Canister Assembly up and out of the Base Cabinet.
10. Remove the Blow Down Muffler from the Sieve Canister Assembly. Refer to the Blowdown Muffler
Replacement Section.
11. Remove the brass fitting from the top of the Sieve Canister Assembly.
To install the Sieve Canister Assembly
1. Install the supplied canister gasket on the bottom of the Sieve Canister Assembly.
2. Install the brass fitting in the top of the Sieve Canister Assembly.
3. Install the Blow Down Muffler onto the Sieve Canister Assembly Refer to the Blowdown Muffler
Replacement Section.
4. Set the Sieve Canister Assembly in place in the Base Cabinet.
5. Slide the one-eared clamp provided onto the 12” pressure tubing.
6. Install the 12” pressure tubing provided on the inlet port of the Sieve Canister.
7. Using a crimping clamp tool or diagonals, “pinch” the one-eared clamp securing the pressure tub-
ing to the inlet port of the Sieve Canister Assembly.
8. Using a phillips screwdriver, insert then tighten the four screws that secure the Sieve Canister
Assembly to the molded base.
9. Install the Valve/Solenoid on the top of the Sieve Canister.
10. Install the wire connectors on the terminals of the valve/solenoid.
11. Connect the yellow pressure tubing to the brass fitting on the top of the Sieve Canister.
12. Connect the clear pressure tubing to the port on the pressure regulator.
NOTE
Ensure that the AC Power Cord is clear of the Sieve Canister
Assembly.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...