7. Monitor the pressure gauge. The gauge should cycle as listed below:
•
H-600/605 120V Concentrators between 25 and 29 psi @ 5 lpm.
•
H600/605 230V Concentrators between 21 and 25 psi @ 5 lpm.
•
M600/605; Enhanced M600/605 120V Concentrators between 25 and 29 psi @ 5 lpm.
•
M600/605; Enhanced M600/605 230V Concentrators >15 psi @ 5 lpm.
8. Read the peak pressure for four cycles, and confirm that all peaks are within one (1) psig of each
other.
9. Disconnect the pressure gauge, “T” fitting, and tubing. Reconnect the yellow tubing to the fitting
on the top of the Sieve Canister.
10. Install the Front and Rear Cabinets.
O
XYGEN
O
UTLET
/R
EGULATED
P
RESSURE
T
EST
1. Turn on the unit by moving the Power Switch to the ON (I) position. Let the unit run for at least two
(2) minutes to stabilize.
2. Set the Flow Meter to 5 lpm (1 lpm for concentrators equipped with a Pediatric Flow Meter).
3. Connect the pressure gauge from the tool kit to the DISS outlet fitting. The flow ball should drop to
0.0 lpm.
4. If the oxygen pressure does not fall between 5.0 and 7.0 psig, perform steps 5 through 13. If the
pressure does fall between 5.0 and 7.0 psig proceed to next test.
5. Remove the pressure gauge from the DISS outlet fitting.
6. Remove the Front and Rear Cabinets.
7. Insert a “T” from the tool kit in between the regulator and the clear oxygen tubing, and connect the
pressure gauge.
8. Set the Flow Meter to 5 lpm (1 lpm for concentrators equipped with a Pediatric Flow Meter).
9. With the gauge held in a vertical position, verify that the regulated pressure is at 5.5 ± 0.25 psig.
NOTE
If the pressure is not within specifications and there are no leaks in the system, perform
Compressor maintenance.
NOTE
If the four cycles are not within specification, proceed to the system pressure test table, located
in the Troubleshooting Section, for diagnostic information.
NOTE
When testing a concentrator equipped with a pediatric Flow Meter, substitute 1 lpm
where 5 lpm is called out in this test.
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...