
T
ESTING
- P
AGE
7
1041900, R
EV
. 00
M
ILLENNIUM
S
ERVICE
& T
ECHNICAL
I
NFORMATION
M
ODEL
M10605 T
EST
D
ATA
S
HEET
NOTE
All information on this data sheet should be entered in the correct location after the associated test was
completed. The data sheet must then be signed in ink and dated by the technician performing tests.
Step 2
Serial Number: __________
Step 3
VAC: __________ Hz:__________
Step 4
Hour Meter: __________
Step 5
No Flow Alarms
< 60 sec.
PASS
FAIL
Step 6
Oxygen Putlet Pressure
5.0-7.0 psig
PRESSURE GAUGE READING
Step 7
Oxygen
Analyzer
Reading
1ST CYCLE
2ND CYCLE
3RD CYCLE
Average O2 Purity (3 cycles)
DMM Voltage Reading
PASS
FAIL
Step 8
Flow Meter
Oxygen
Concentration
M10605;
Enhanced M10605
120V
92% +/- 4 @ 10 lpm
94% +/- 2 @ 3 lpm
3 lpm
10 lpm
Signature: ____________________ Date:_______________ Notification Number: _____________
(RI internal use Only)
Содержание Millennium Enhanced M600
Страница 1: ...Millennium TM Oxygen Concentrators Service Technical Information ...
Страница 25: ...1041812 REV 00 PAGE 2 SYSTEM SETUP MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally blank ...
Страница 55: ...1041813 REV 00 PAGE 26 MAINTENANCE MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 305: ...1041901 REV 00 PAGE 50 REPAIR KITS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 317: ...1041903 REV 00 PAGE 12 TESTING MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 333: ...1041904 REV 00 PAGE 6 SCHEMATICS MILLENNIUM SERVICE TECHNICAL INFORMATION This page intentionally left blank ...
Страница 334: ...Reorder Number 1041825 1044572 Rev 00 BRM 5 7 2007 ...
Страница 335: ...Service Technical Information Millennium Oxygen Concentrators ...