SECTION 4 PROGRAMMING IN HYGIENE MODE
4 - 15
" GAS On" to remind user to turn on the
standard calibration gas bottle now. After the
gas bottle is turned on, press
[enter]
key to
continue the standard gas calibration procedure
as described in section 4.7.3.
5. Any time during step 2 or step 3, if the
[menu]
key is
pressed, the data entry operation will be aborted. It is also
assumed that user want to skip the standard gas calibration
procedure and move on to the next menu item. Therefore,
it is very important that all 4 digits are entered ( i.e. press
[enter]
key 4 times ) even if user does not want to change
the standard calibration gas value in step 2.