SECTION 9 TROUBLESHOOTING
9 - 4
Problem
Possible Reasons & Solutions
Can not turn off
unit or
corrupted
characters in
LCD display
Reasons:
Microprocessor hang-up
Solutions:
Turn off then turn on
battery switch
Reload software from PC
Full scale
measurement in
humid
environment
Reasons:
Dirty or wet sensor
Solutions:
Clean and dry sensor
Use water trap disk
to block out moisture
Measurement
max out at
certain level
Reasons:
Dirty lamp/ sensor module
Weak lamp
Solutions:
Clean lamp/ sensor module
Replace new lamp
Calibration
error message
Reasons:
No standard gas input
Low sensitivity to cal gas
Solutions:
Make sure standard gas
flows into inlet probe
Change cal gas