SECTION 2 OPERATION OF MINIRAE PLUS
2 - 16
In Survey Mode, the MiniRAE Plus Professional Unit will allow
users to record up to 50 readings of average or peak gas
concentration over a specified sampling period. In addition, a
time stamp and a site number will also be recorded together
with each reading. The sampling period can be programmed
from 1 second up to 9999 seconds.
Users are allowed to select one of the following two different
datalog options:
1) manual sampling option: in this option, users have to press
[enter]
key to start each sampling period. When a sampling
period is over, the datalogging operation will stop.
2) continuous sampling option: in this option, users press
[enter]
key once to start the first sampling period. At the end
of each sampling period, next sampling period will start
automatically. The datalogging operation will not stop until
another
[enter]
key is pressed.
Note: in either option, the datalogging operation will stop when
[enter]
key is pressed any time during a sampling period.
User is also allowed to choose to datalog the
average
value or
the
peak
value of each sampling period (see section 5.14 for
details).
The site number will automatically advance to next one when a
sampling period is over. Users can use "select site number"
menu in programming task to change to a different site number
(see section 5.4 for details).
Содержание MiniRAE Plus PGM-76
Страница 2: ...RAE Systems Inc 680 West Maude Ave 1 Tel 408 481 4999 Sunnyvale CA 94086 Fax 408 481 4998...
Страница 14: ......
Страница 18: ...SECTION 1 GENERAL INFORMATION 1 4 This page is intentionally left blank...
Страница 56: ...SECTION 3 OPERATION OF ACCESSORIES 3 16 This page is intentionally left blank...
Страница 104: ...SECTION 5 PROGRAMMING IN SURVEY MODE 5 18 This page is intentionally left blank...