SECTION 9 TROUBLESHOOTING
9 - 3
Problem
Possible Reasons & Solutions
Reading jump
around
randomly
Reasons:
Incorrect gas calibration
Low sensitivity to cal gas
Solutions:
Re-calibrate
Use different cal gas
Slow response
to gas input
Reasons:
Leakage in probe assembly
or sensor module
Solutions:
Tighten the probe
assembly and sensor
module
No air draw at
gas inlet tube
Reasons:
Defective pump or leakage
in probe assembly and
sensor module
Solutions:
Replace pump, tighten the
probe assembly and
sensor module
"Lo bAt"
message at
power on
Reasons:
Discharged battery
Solutions:
Recharge battery