SECTION 4 PROGRAMMING IN HYGIENE MODE
4 - 28
4.11 Set Security Level
In Hygiene Mode, authorized users are allowed to set the
security level from the PC so that three levels of different data
protection can be setup for the operator of the instrument.
1) level 0 security: in this level, users can view three alarm limits
and the current time and date in programming task. They
are not allowed to make any changes.
2) level 1 security: in this level, users have to enter a 4 digits
password before they can enter the programming tasks.
Once in the programming task, they are allowed to change
all menu items.
3) level 2 security: in this level, users are allowed to make any
changes in programming task. There is no need for a
password to enter the programming task.
The security level has to be set up by "ProRAE-76" software in
PC. See section 6 for details on how to install the ProRAE-76
software on PC, how to interface the MiniRAE Plus instrument
to PC, and how to set up security levels in PC.
A brief summary on how to set up security level in PC includes:
Step 1: From the main menu of the ProRAE software,
Open
the Hygiene Mode configuration (hygiene.CFG ) file from the