SECTION 4 PROGRAMMING IN HYGIENE MODE
4 - 12
Zero Gas Calibration Procedure
1. "Zero gas calibration" option is the 5th menu option in Table
4.1.
- Display shows "C0 xxx.x" where
"xxx.x" is the gas reading based on current
calibration of the instrument. Note: this
reading may not be zero due to back ground gas
concentration, dirty charcoal filter or instrument
drift .
2. Insert the zero gas adapter into the gas inlet tube of the
Personal PID Unit, as shown in Fig. 4.1.
3. The sampling pump draws air sample into the sensor
chamber through the charcoal filter of the organic zeroing
kit. The air entering the sensor chamber should be free of
any organic vapors.
- Display then shows the reading of the gas free
of organic vapor. Again, this reading may not
be zero due to dirty charcoal filter or instrument
drift .