SECTION 2 OPERATION OF MINIRAE PLUS
2 - 14
2.7 Preset Alarm Limits and
Calibration
The MiniRAE Plus Professional PID Unit is factory calibrated
with standard isobutylene gas, and is programmed with default
alarm limits as listed below. Refer to section 4 and 5 for
programming procedures if new calibration or new alarm limits
are required.
Table 2.2
Factory Calibration and Preset Alarm Limits
Calibration Standard Gas
100 ppm isobutylene
Hygiene Mode:
TWA Alarm Limit
10 ppm
STEL Alarm Limit
25 ppm
Peak Alarm Limit
50 ppm
Survey Mode:
Low Alarm Limit
10 ppm
High
Alarm Limit
100 ppm
Battery Low Level
5.5 volt
Power Off Level
5.3 volt
Содержание MiniRAE Plus PGM-76
Страница 2: ...RAE Systems Inc 680 West Maude Ave 1 Tel 408 481 4999 Sunnyvale CA 94086 Fax 408 481 4998...
Страница 14: ......
Страница 18: ...SECTION 1 GENERAL INFORMATION 1 4 This page is intentionally left blank...
Страница 56: ...SECTION 3 OPERATION OF ACCESSORIES 3 16 This page is intentionally left blank...
Страница 104: ...SECTION 5 PROGRAMMING IN SURVEY MODE 5 18 This page is intentionally left blank...