SECTION 2 OPERATION OF MINIRAE PLUS
2 - 21
2.12 Diagnostic Power-on Sequence
The MiniRAE Plus Professional PID Unit offers a special
diagnostic power on sequence to allow user to read "raw" (or
un-calibrated) instrument readings during zero gas calibration
procedure. This feature can provides additional information to
users regarding the sensor performance. This is accomplished
by pressing
[up]
key first, holding it down and then pressing
[on]
key to turn on power.
Note:
using the special diagnostic power on sequence, the unit
will perform normally except during zero gas calibration
procedure ( see section 4.7.1). The display will show a "raw"
reading, instead of the "ppm" reading. Typically, the reading
should be between "330.0" to "370.0". When exposed to a
calibration gas, the reading should go down quickly. The
reading can be anywhere between "0.0" to "300.0".
If the reading does not change when exposed to a calibration
gas, then it is an indication that the lamp is very weak or the
sensor module is defective.