- 51 -
13.2. Adjustment menu settings
P1 CALIBRATION
P1.1
INT.CALIB.
|
[internal
adjustment]
P1.2
EXT.CALIB.
|
[external
adjustment]
P1.3
USER CALIB.
|
[user adjustment]
P1.4
CALIB. TEST
|
[adjustment test]
P1.5
AUTO. CAL. T.
| NONE/TIME/TEMP./BOTH
[automatic
adjustment]
P1.6
AUTO CAL. C.
|
[time of
automatic
adjustment]
•
P1.1
INT. CALIB.
– Internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically,
and no user assistance is required. If a weighing pan of the balance is loaded,
balance will display a command to unload the weighing pan. If weighing pan is
loaded with a relatively small mass (± 10% of balance max capacity) then
adjustment process will be carried out automatically, no user ingerention
needed.
or
•
P1.2
EXT. CALIB.
– External adjustment (with the use of an
external weight)
Adjustment with an external weight, value of which is saved in factory settings,
function unavailable for verified balances.