- 106 -
17.3.2.
Setting the reference mass by determining it from a sample
of specified quantity
While determining the mass of a single piece
ACAI
function (Automatic
Accuracy Correction) is in use.
Means of operation of ACAI function:
−
Number of pieces (on adding) on balance weighing pan has to be
greater than before
−
Number of pieces (on adding) on balance weighing pan must be less
than twice the amount displayed before adding parts
−
Current quantity of parts must be within the
±
0,3 tolerance of the total
value,
−
Measurement result has to be stabilised.
Procedure:
•
Place the container on the pan and tare its mass,
•
Press one F button to which the function <ENTER SAMPLE> is
assigned, wait to see the editing window <SAMPLE QUANTITY>
•
Use navigating buttons
or
to select the correct sample
quantity.
•
For option: sample quantity (displayed value <0000>) enter any numer
using navigating buttons.