- 173 -
Response format:
1 2-3
4
5
6
7-15 16
17 18 19 20 21
S space stability
marker space character mass space unit
CR LF
Example
S CR LF
– command sent from a computer
S _ A CR LF
– command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
– command carried out,
response is weight value in basic measuring unit.
Immediately send measurement result in basic measuring unit
Format:
SI CR LF
Accessible responses:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response is immediate with weight value in basic
weighing unit
Response format:
1 2 3
4
5
6
7-15 16
17 18 19 20 21
S I space stability
marker space character mass space unit
CR LF
Example
S I CR LF
– command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
– command carried out, immediate
response of weight value in a basic weighing unit.
Send stable measurement result in current measuring unit
Format:
SU CR LF
Accessible responses:
SU_A CR LF
SU_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable
measurement result
SU_I CR LF
- command understood but not accessible at this
moment
SU_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Response format:
1 2 3
4
5
6
7-15 16
17 18 19 20 21
S U space stability
marker space character mass space unit
CR LF