- 54 -
13.3.2.
External adjustment
The external adjustment for PS series balances should be carried out with an
external mass standard / weigh of class F
1
.
Run an external adjustment process,
Balance displays a command ordering to unload the weighing pan,
<REMOVE
MASS>
(the weighing pan must be empty). When the weighing pan is
unloaded, press
button. The balance determines mass of an empty pan,
message <
CALIBRATION
> is displayed in the bottom line. Next, message
<
PLACE MASS
> and mass value to be placed on the weighing pan are
displayed,
e.g.
200.000g
(depending on the type of balance). Place an external
adjustment weight of displayed mass value and press
button
.
The
balance determines the mass, message <
CALIBRATION
> is displayed in the
bottom line. On completing adjustment process the balance returns to submenu
P1.2 EXT.CALIB.
If a balance is verified, it is not possible for a user to carry out the
external adjustment process.
13.3.3.
User adjustment
The external adjustment for PS series balances should be carried out with an
external mass standard / weigh of class F
1
.
Run an external adjustment process