9.4. WEIGHING mode - settings
The software allows setup of operating parameters (filters, value release and
autozero option, deleting the last digit and other settings) separately for each
working mode.
It enables customising the instrument and utilising its properties depending on
the user’s needs and expectations, or on specific requirements for selected
working mode (e.g. DOSING); as a result the device operation is quick and
easy.
Available settings:
P2.2.1 – READING
P2.2.1.1 – FILTER [V.FAST/ FAST / NORMAL / SLOW / V. SLOW]
P2.2.1.2 – RESULT CONF. [REL. / FAST / FAST+REL.]
P2.2.1.3 – AUTOZERO [YES / NO]
P2.2.1.4 – LAST DIGIT [ALWAYS / NEVER / WHEN STAB.]
P2.2.1.5 – AMBIENT CONDITIONS (STABLE/UNSTABLE)
P2.2.2 – AUTOTARE [YES / NO]
P2.2.3 – PRINT MODE
P2.2.3.1 – MODE [EACH / WHEN STAB. / AUTO]
P2.2.3.2 – AUTO. PROG
P2.2.4 – INFORMATION [TARE / NETTO / GROSS / USER / PRODUCT /
/NSTD. INF. / NONE]
P2.2.5 – INF. NSTD. (non-standard information – any text, max 19 characters)
P2.2.6 – HOT KEYS (available options depending on the working mode)
P2.2.6 1 – F1 [NONE / ENTER TARE/ …]
P2.2.6.2 – F2 [NONE / ENTER TARE / …]
P2.2.6.3 – F3 [NONE / ENTER TARE/ …]
P2.2.6.4 – F4 [NONE / ENTER TARE / …]
9.4.1.
Filter level setting
Filter shall be set depending on the ambient and environment conditions. For
the best possible conditions the filter can work in a very fast mode (V.FAST
value for parameter P.2.2.1.1 Filter); however, if the conditions are bad (shakes,
drafts), the filter should be set to slow or very slow option (SLOW or V. SLOW
value for parameter P.2.2.1.1 Filter). The effectiveness of the filter is different
for the weighing range. The filter works slower when “approaching” the weighed
mass, it works more intensively for weighed mass within the set range of the
filter (parameter for setting filter range is accessible only from the service menu
– the user does not have an access to it).
Depending on the filter, the weighing time is shorter (V.FAST and FAST) or
longer (SLOW and V. SLOW).
Содержание PS 3000.R2
Страница 2: ...2 FEBRUARY 2015...
Страница 12: ...12 A Install an assembly with an openwork pan onto the weighing pan base...
Страница 103: ...103 17 2 7 Return to the weighing mode...
Страница 120: ...120 17 6 3 Return to the weighing mode...
Страница 141: ...141 17 10 3 Return to the weighing mode...