- 158 -
Accessible options:
COM 1
or
COM 2
– RS 232 port, to which the computer is connected
USB
type B – USB port, to which the computer is connected
WIFI
– WIFI port
19.1.2.
Continuous transmission
Continuous transmission parameter enables turning on/off continuous
transmission of a measurement.
It is possible to set one of three available options:
−
Continuous transmission in a calibration unit <CAL. UNIT>,
independently from currently chosen weight unit, measurements are
transmitted in unit set on a main measuring device (calibration unit),
−
Continuous transmission in a current unit <CURRENT UNIT>
measurements are transmitted in a current unit; the unit changes
dynamically whenever the user presses Units button in order to modify
the unit on a display,
−
Continuous transmission off, <NONE>.
Setting instruction:
Caution:
Continous transmission may be turned on/off by means of command sent from
a computer (go to COMMUNICATION PROTOCOL section).