- 167 -
20.3.
Format of data sent on pressing PRINT button
CAUTION
Unstable measurement printout is not available for a verified
balance.
Printout format
1
2
3
4 - 12
13
14 - 16
17
18
st
ab
ilit
y
m
ar
ker
sp
ac
e
char
ac
ter
m
as
s
sp
ac
e
uni
t
CR
LF
stability marker - [space] – when stable
- [?] – when unstable
- [^] – if maximum range is exceeded
- [v] – if minimum range is exceeded
character
- [space] – for positive or [-] for negative values
mass
- 9 characters – aligned to the right
unit
- 3 characters – aligned to the left.
20.3.1.
Format of data sent as a response for commands generated
by a computer
After receiving a command, balance responds as follows first:
XX_A CR LF
command understood and in progress
XX_I
CR LF
command understood but not accessible at this moment
XX_^ CR LF
command understood but maximum range is exceeded
XX_v CR LF
command understood but minimum range is exceeded
XX_E CR LF
an error occurred on command carrying out (time limit
exceeded while waiting for stable measurement result),
time limit is balance’s characteristic parameter
XX
name of the command
Содержание PS 3000.R2