- 32 -
For correct operation of the balance with AUTOTARE function, the threshold
value must be adjusted.
<P 2.2.3.2 AUTO THRES >
parameter is connected with the following functions:
-
automatic tare,
-
automatic operation,
There will be no automatic tarring until the balance operates within the gross
weight value range set in
< P 2.2.3.2 AUTO THRES >
parameter.
9.6. Printing mode
This function enables setting the printing mode, i.e.
button activity.
A user has the possibility of selecting the following settings:
•
<WHEN STAB> the stable measurement result, along with the settings
for parameter <GLP PRINTOUT> are sent to the printer port. On
pressing
button, when the result is not stable (no
sign on
a display), the balance software will send the measurement result to
the port after reaching stability for the measurement.
•
<EACH> every single pressing of
button results in sending the
measurement result to the printer port along with the settings for
parameter <GLP PRINTOUT>. Every measurement result is sent
(stable and unstable). The <?> sign appears at the beginning of the
printing
frame
when
the
result
is
unstable.
This function applies only to non-verified balances.
•
<AUTO> - select this option to enable automatic printing of
measurements. If this option has been selected, remember to select
<AUTO THRES > parameter to suit your needs.
Automatic operation procedure:
•
Press
button to zero the balance (marker of stable
measurement
and zero marker
are shown on a display)
•
Deposit load, the balance sends the first stable measurement to the
printer port,
Содержание PS 3000.R2
Страница 2: ...2 FEBRUARY 2015...
Страница 12: ...12 A Install an assembly with an openwork pan onto the weighing pan base...
Страница 103: ...103 17 2 7 Return to the weighing mode...
Страница 120: ...120 17 6 3 Return to the weighing mode...