An exemplary report:
.....................................................
--------- Autotest GLP: Report ---------
Balance type PS 3000.R2
Balance ID 400010
User Admin
Software rev. v.0.4.9
Date 2013.07.16
Time 09:17:16
----------------------------------------
Number of measurements 10
Reading unit 0.001/0.01 g
Internal weight mass 1402.094 g
Filter Average
Value release Fast &Reliable
----------------------------------------
Deviation for Max. -0.118 g
Repeatability 0.0088 g
Signature
........................................
Procedure:
Parameter <P6.11 GLP AUTOTEST > is initiated by pressing
button.
The balance software starts the autotest GLP procedure and it is carried out
automatically from now on to the end.
A user can stop the procedure at any time by pressing
button of the GLP
process window.
Once the procedure is finished, the value of standard deviation of all
measurements is shown in the main display, and message <RESULT> in the
bottom line, the final report is printed out automatically (as shown above). It is
possible to print out the report again by pressing
button.
Report is not saved, therefore when abandoning this level, it is deleted from the
balance volatile memory.
To abandon the window press
button and the program returns to the main
menu. To return to weighing press
button repeatedly.