Example
S U CR LF
– command sent from a computer
S U _ A CR LF
– command understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF
– command carried out
response is mass value in current measuring unit.
Immediately send measurement result in current measuring unit
Format:
SUI CR LF
Accessible answers:
SUI_I CR LF
- command understood but not accessible at this moment
MASS FRAME - mass value in current measuring unit is returned
immediately
Response format
1 2 3 4
5
6
7-15 16
17 18 19 20 21
S U I stability
marker space character mass space unit
CR LF
Example
S U I CR LF
– command from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
– command carried out, immediate
response of mass value in current measuring unit
where: _ - space
Switch on continuous transmission in basic measuring unit
Format:
C1 CR LF
Accessible answers:
C1_I CR LF
- command understood but not accessible at this
moment
C1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Response format:
1 2 3
4
5
6
7-15 16
17 18 19 20 21
S I space stability
marker space character mass space unit
CR LF
Switch off continuous transmission in basic measuring unit
Format:
C0 CR LF
Accessible responses:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out