- 45 -
11.10. Time format
Parameter which enables choosing time format on the printout [12H / 24H].
For the value [
12H
], set in parameter
P6.6 TIME,
letters
<
A
> or <
P
> will be
displayed next to the time value.
Where:
A
stands for hours before noon,
P
stands for hours after noon.
Letters
AM
or
PM
are printed out next to the time value.
11.11. Autotest GLP
AUTOTEST function is designed to aid a user in assessing balance’s operation
and diagnosing the reasons for occurrence of errors in weighing which exceed
the maximum permissible values for a given balance model.
By means of a simple, repeatable and fully documented way, the function
enables optimizing balance’s settings to maintain the best possible repeatability
and weighing time at workstation. The main purpose of the function is the
possibility of monitoring the above-mentioned parameters at optional moment
and saving records from the carried out tests in the form of printed reports of the
tests that are automatically generated at the end of examination.
The test controls repeatability of placing the interval weight and determining
error of indication with reference to balance’s maximum capacity.
Testing procedure:
•
deposit internal weight twice,
•
deposit internal weight ten times,
•
perform balance adjustment,
•
calculate the value of standard deviation,
•
print a report.
Test results:
*Deviation for Max.
*Value of repeatability of indication expressed as standard deviation.